How much motion makes a chest ECG unusable?
Proposed study of movement artefacts, missing samples and whether a quality flag improves an interval display. Inside AI has collected no recordings.
Device profile · Movesense HR+ Developer Kit ↗Collect comparable segments
Record ECG and acceleration concurrently with the same phone, firmware, sample rates, sensor and placement. Label short sessions at rest and during walking. Separately mark observed strap contact changes. Record the activity and contact condition independently, since motion alone does not prove poor ECG quality.
Score the decision
Have a reviewer label held-out windows for beat visibility, including an uncertain class. Compare a simple ECG-only quality rule with a model that also uses acceleration. Hold out whole sessions to reduce leakage. Report how many poor windows each method accepts, how often it suppresses usable windows, and how much time remains eligible for a heart-rate variability (HRV) display. Show a gap when quality is insufficient.
Decide where computation belongs
Check sample gaps and timestamp offsets after reconnecting. If a quality flag is repeatable on the host, try a separate embedded motion-event detector using the Device API. Measure free RAM, binary size and current draw under matching sample rates before choosing sensor-side processing. No accuracy, endurance or on-device fit result is available.